Max-550E Oxygen Sensor for MaxBlend

Replacement For

  • Maxtec [MaxBlend 2, MaxBlend Lite, BlenderBuddy 2, R140P02-001]
  • CareFusion [Sentry 2, R140P02-001]

Specifications

  • Range of Measurement: 0 to 100%
  • Output Range: 10.5 to 15 mV
  • Zero Offset: ≤ 0.50 mV
  • Response Time: ≤ 15 s
  • Linearity: < 2.0%
  • Expected Lifetime: 1,500,000 %O2 hours
  • Temperature Compensation: NTC
  • Operational Temperature Range: 15° to 40°C
  • Storage Temperature Range: -15° to 50°C
  • Humidity Range: 0% to 95% RH non-condensing
  • Ambient Pressure Range: 800-1013 mbar
  • Interference: Per EN ISO 80601-2-55
  • Electrical Interface: 3-pin stereo jack

Request More Information