Max-250TM Oxygen Sensor

Replacement For

  • Medtronic [PB-840, 980]
  • Custom [Custom applications with 3-pin molex connector]

Specifications

  • Range of Measurement: 0 to 100% O2
  • Output Range: 11 ± 2.5 mV
  • Zero Offset: ≤ 0.50 mV
  • Response Time: ≤ 15 s
  • Linearity: ± 2.0%
  • Expected Lifetime: > 1,500,000 %O2 hours
  • Temperature Compensation: < ± 3.0%
  • Operational Temperature Range: 5˚ to 40˚C
  • Storage Temperature Range: -15˚ to 50˚C
  • Humidity Range: 5% to 95% RH, non-condensing
  • Interference: Per EN ISO 80601-2-55