Max-14 Oxygen Sensor

Replacement For

  • Devilbiss [02ANA-101 Sensor]
  • MSA [806572, MiniOX 1A]
  • Mercury Medical [10-103-13]
  • Pacifitech [PT-26]

Specifications

  • Range of Measurement: 0 to 100% O2
  • Output Range: 9.0 to 13.0 mV
  • Zero Offset: ≤ 0.20 mV
  • Response Time: ≤ 12 s
  • Linearity: < 3%
  • Expected Lifetime: > 1,000,000 %O2 hours
  • Operational Temperature Range: 0˚ to 50˚C
  • Storage Temperature Range: -20˚ to 50˚C
  • Humidity Range: up to 99% RH
  • Ambient Pressure Range: 600-1750 mbar
  • Interference: per EN ISO 80601-2-55
  • Electrical Interface: 3-pin Molex