Max-12a Oxygen Sensor

Replacement For

  • HillRom [Isolette Incubator]
  • Philips (Respironics) [NeoPAP]

Specifications

  • Range of Measurement: 0 to 100%
  • Output Range: 8.0 to 13.0 mV
  • Zero Offset: ≤ 0.200 mV
  • Response Time: ≤ 15 s
  • Linearity: ≤ 3%
  • Expected Lifetime: > 500,000 %O2 hours
  • Temperature Compensation: NTC
  • Operational Temperature Range: 10° to 40°C
  • Storage Temperature Range: -20° to 50°C
  • Humidity Range: 0% to 99% RH, non-condensing
  • Ambient Pressure Range: 700-1250 mbar
  • Interference: PER EN ISO 80601-2-55
  • Electrical Interface: 3-pin Molex

Request More Information